Измеритель коэффициента экстинкции N7782B
Alignment of polarization-maintaining fiber optic devices requires a PER-analyzer. Keysight’s N7782B has been designed for high speed and highly accurate testing of the polarization extinction ratio (PER) in PM fibers and devices.
Характеристики
Производитель
—
Keysight Technologies
Модель
—
N7782B
Измеритель коэффициента экстинкции N7782B
HIGHLIGHTS
Key Benefits
Key Benefits
- Accurate polarization extinction ratio measurements in PMF, up to 50 dB
- Uses the full polarimetric method recommended for aligning PMF with laser signals
- Supports automated test with N7783B thermal cycling units for fiber stressing
- Easy to use graphic software with real-time display
- Laser Diode PMF Pig-tailing Alignment of the PM fiber during the pig-tailing process is supported by real-time display of the PER and the optical power.
- PMF Splicing In order to support the alignment during the splicing process of PM fibers the Keysight N7782B provides real time display of the optical power and of the angular misalignment of the two fibers
- PM component Characterization measurement of the PER on PM components like fiber polarizers, PMF couplers, PMF splitters, etc.
- Characterization of PMF cross-coupling Polarization crosstalk in a PM fiber is measured and displayed as PER
The real time measurement capability in combination with automation interfaces makes this unit ideally suited for integration in manufacturing systems, for example pig-tailing stations for laser diodes and planar waveguide components. Analog interfaces are provided for integration of the system in control loop applications.
Документы
Wavelength | ||
Specification wavelength range | O-Band (Opt. 300) | 1270 nm to 1375 nm |
O/C/L-Band (Opt. 400) | 1270 nm to 1375 nm | |
1460 nm to 1620 nm | ||
C/L-Band (Opt. 500) | 1460 nm to 1620 nm | |
Operating wavelength range 2 | 1260 nm to 1640 nm | |
PER analysis | ||
PER range 3, 4 | 0 dB to 50 dB | |
PER uncertainty, single-TCU method (typical) 3, 4 |
PER = 0 dB to 30 dB | 0.30 dB |
PER = 30 dB to 50 dB | 0.60 dB | |
Splice angle analysis | ||
Splice angle uncertainty, dual-TCU method (typical) 3, 4 | ± (0.1° + 4% x angle) | |
Optical power | ||
Input power range | –50 dBm to +7 dBm | |
Relative power uncertainty 3 | C/L-band: ± 30 mdB (±20 mdB, typ.) | |
O-band: ± 70 mdB (± 40 mdB, typ.) | ||
Internal laser source | ||
Wavelength | O-Band (Opt. 401) | 1290 nm to 1360 nm |
1310 nm typical | ||
C-Band (Opt. 501, 401) | 1510 nm to 1580 nm | |
1550 nm typical | ||
Output power (typical) 5 | O-Band (Opt. 401) | –12 dBm |
C-Band (Opt. 501, 401) | –10 dBm |
1.
Ambient temperature change max. ± 0.5 °C since normalization. Specification valid on day of calibration.
2. PER measurements are possible outside the specification wavelength range if the user performs a manual calibration. Note that a fully polarized light source is needed for calibration.3. Input power > –30 dBm.
4. Narrow-band light source with DOP > 95% needed.
5. At room temperature.