Векторный анализатор цепей E5072A
Versatile, fast, and flexible
Характеристики
Производитель
—
Keysight Technologies
Модель
—
E5072A
Векторный анализатор цепей E5072A
Designed for the Most Demanding Requirements from Research and Development to Manufacturing
The Keysight E5072A vector network analyzer, with a configurable test set, provides access to the signal paths between the internal source, receivers, bridges, and the analyzer’s test ports. The direct access improves instrument sensitivity as well as the ability to add components or peripherals for a variety of applications. The E5072A delivers new standards in versatility, speed, and accuracy, and is suitable for full performance characterization of passive and active components such as antennas, filters, and high-power amplifiers.
- Choose frequency from 30 kHz to 4.5 GHz or 8.5 GHz; 2 ports
- Improve accuracy, yield, and margins with a wide dynamic range of 130 dB / 151 dB with direct receiver access
- Increase throughput with measurement speeds of 7 ms at full 2-port calibration, 401 points
- Reduce calibration time with excellent temperature stability of 0.005 dB/°C
- Gain test flexibility with a wide source power range of -109 to +20 dBm for linear and nonlinear device characterization
- Easy to configure test set provides access to the signal paths between the internal source, receivers, bridges, and the analyzer’s test ports.
High-power measurement
High-power measurements are necessary to characterize devices under real conditions. These measurements are cumbersome if external amplifiers and attenuators are attached to a network analyzer's test ports. The E5072A reduces the need for external power sources by delivering up to +20 dBm from its test ports. For measurements above +20 dBm, the E5072A's configurable test set provides direct source and receiver paths for simple test setup and error correction for external devices.
Документы
Corrected System Performance
The specifications in this section apply to measurements made with the Keysight Technologies, Inc. E5072A network analyzer
under the following conditions:
1. The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The
effective dynamic range must take measurement uncertainty and interfering signals into account.
2. Typical performance might not be met from 60 to 70 kHz.
3. The direct receiver access input system dynamic range is calculated as the difference between the direct receiver access input noise floor and the
source maximum output power. The effective dynamic range must take measurement uncertainties and interfering signals into account.
Test Port Input
Trace Noise 2
1. Typical performance might not be met from 60 to 70 kHz.
2. The specification might not be met at the following frequencies: 333.333 kHz, 406.25 kHz, 857.143 kHz, and 928.571 kHz.
Stability
Dynamic Accuracy 1
1. Accuracy of the test port input power reading is relative to –10 dBm reference input power level.

The specifications in this section apply to measurements made with the Keysight Technologies, Inc. E5072A network analyzer
under the following conditions:
- No averaging applied to data
- Environmental temperature of 23 °C (± 5 °C) with less than 1 °C deviation from the calibration temperature
- Response and isolation calibration performed
Description | Specification | SPD |
System dynamic range at test port 1, 2 | ||
(IF Bandwidth = 3 kHz) | ||
30 to 300 kHz | 65 dB (Typ.) | |
300 kHz to 10 MHz | 82 dB | |
10 MHz to 6 GHz | 98 dB | |
6 to 8.5 GHz | 92 dB | |
(IF Bandwidth = 10 Hz) | ||
30 to 300 kHz | 90 dB (Typ.) | 100 dB |
300 kHz to 10 MHz | 107 dB | 115 dB |
10 MHz to 6 GHz | 123 dB | 130 dB |
6 to 8.5 GHz | 117 dB | 128 dB |
Extended dynamic range at direct receiver access input 3 | ||
(IF Bandwidth = 10 Hz) | ||
30 to 300 kHz | 128 dB | |
300 kHz to 10 MHz | 134 dB | |
10 MHz to 3 GHz | 151 dB | |
3 GHz to 6 GHz | 147 dB | |
6 to 8.5 GHz | 145 dB |
1. The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The
effective dynamic range must take measurement uncertainty and interfering signals into account.
2. Typical performance might not be met from 60 to 70 kHz.
3. The direct receiver access input system dynamic range is calculated as the difference between the direct receiver access input noise floor and the
source maximum output power. The effective dynamic range must take measurement uncertainties and interfering signals into account.
Test Port Input
Description | Specification | SPD |
Crosstalk 1 30 to 300 kHz 300 kHz to 10 MHz 10 MHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz |
–90 dB (Typ.) –110 dB –120 dB –110 dB –100 dB |
–95 dB –120 dB –140 dB –130 dB –120 dB |
Test Port Noise Floor 1 30 to 300 kHz 300 kHz to 10 MHz 10 MHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz |
–90 dBm/Hz (Typ.) –101 dBm/Hz –117 dBm/Hz –121 dBm/Hz –117 dBm/Hz |
–100 dBm/Hz –109 dBm/Hz –124 dBm/Hz –128 dBm/Hz –128 dBm/Hz |
(at 10Hz IFBW) 30 to 300 kHz 300 kHz to 10 MHz 10 MHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz |
–80 dBm (Typ.) –91 dBm –107 dBm –111 dBm –107 dBm |
–90 dBm –99 dBm –114 dBm –118 dBm –118 dBm |
Compression level (at +10dBm input) Magnitude 30 to 300 kHz 300 kHz to 3 GHz 3 to 8.5 GHz Phase 30 to 300 kHz 300 kHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz |
0.3 dB (Typ.) 0.2 dB 0.2 dB 7 deg (Typ.) 5 deg 5 deg 5 deg |
0.15 dB 0.08 dB 0.1 dB 0.7 deg 0.3 deg 0.6 deg 1.0 deg |
0.1 dB Compression Input Level 30 to 300 kHz 300 kHz to 2 GHz 2 to 6 GHz 6 to 8.5 GHz |
6 dBm 16 dBm 14 dBm 10 dBm |
Trace Noise 2
Description |
Specification |
SPD |
(test port input level = maximum power in Specification) |
||
Magnitude 30 to 300 kHz, 3 kHz IFBW 300 kHz to 10 MHz, 3 kHz IFBW 10 MHz to 4.38 GHz, 70 kHz IFBW 4.38 to 8.5 GHz, 70 kHz IFBW |
0.015 dB rms (Typ.) 0.003 dB rms 0.004 dB rms 0.006 dB rms |
0.007 dB rms 0.0005 dB rms 0.001 dB rms 0.0012 dB rms |
Phase 30 to 300 kHz, 3 kHz IFBW 300 kHz to 10 MHz, 3 kHz IFBW 10 MHz to 4.38 GHz, 70 kHz IFBW 4.38 to 8.5 GHz, 70 kHz IFBW |
0.1 deg rms (Typ.) 0.02 deg rms 0.035 deg rms 0.05 deg rms |
0.05 deg rms 0.0025 deg rms 0.0075 deg rms 0.015 deg rms |
1. Typical performance might not be met from 60 to 70 kHz.
2. The specification might not be met at the following frequencies: 333.333 kHz, 406.25 kHz, 857.143 kHz, and 928.571 kHz.
Stability
Description | Specification | SPD |
Magnitude 30 to 300 kHz 300 kHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz |
± 0.04 dB/°C ± 0.005 dB/°C ± 0.03 dB/°C ± 0.04 dB/°C |
|
Phase 30 to 300 kHz 300 kHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz |
± 0.8 deg/°C ± 0.1 deg/°C ± 0.4 deg/°C ± 0.8 deg/°C |
Dynamic Accuracy 1
Description | Specification (dB) | Typical |
Magnitude 10 dBm –30 dBm –100 dBm –110 dBm |
± 0.21 dB ± 0.045 dB ± 2 dB |
± 3.0 dB |
Phase 10 dBm –30 dBm –100 dBm |
± 5 deg ± 0.3 deg ± 15 deg |